Skip to main content
Close
Publicació

A Follow-Up of the Multicenter Collaborative Study on HIV-1 Drug Resistance and Tropism Testing Using 454 Ultra Deep Pyrosequencing.

Veure totes les publicacions

Background: Ultra deep sequencing is of increasing use not only in research but also in diagnostics. For implementation of ultra deep sequencing assays in clinical laboratories for routine diagnostics, intra- and inter-laboratory testing are of the utmost importance.

Methods: A multicenter study was conducted to validate an updated assay design for 454 Life Sciences' GS FLX Titanium system targeting protease/reverse transcriptase (RTP) and env (V3) regions to identify HIV-1 drug-resistance mutations and determine co-receptor use with high sensitivity. The study included 30 HIV-1 subtype B and 6 subtype non-B samples with viral titers (VT) of 3,940-447,400 copies/mL, two dilution series (52,129-1,340 and 25,130-734 copies/mL), and triplicate samples. Amplicons spanning PR codons 10-99, RT codons 1-251 and the entire V3 region were generated using barcoded primers. Analysis was performed using the GS Amplicon Variant Analyzer and geno2pheno for tropism. For comparison, population sequencing was performed using the ViroSeq HIV-1 genotyping system.

Results: The median sequencing depth across the 11 sites was 1,829 reads per position for RTP (IQR 592-3,488) and 2,410 for V3 (IQR 786-3,695). 10 preselected drug resistant variants were measured across sites and showed high inter-laboratory correlation across all sites with data (P0.001). The triplicate samples of a plasmid mixture confirmed the high inter-laboratory consistency (mean% ± stdev: 4.6 ±0.5, 4.8 ±0.4, 4.9 ±0.3) and revealed good intra-laboratory consistency (mean% range ± stdev range: 4.2-5.2 ± 0.04-0.65). In the two dilutions series, no variants >20% were missed, variants 2-10% were detected at most sites (even at low VT), and variants 1-2% were detected by some sites. All mutations detected by population sequencing were also detected by UDS.

Conclusion: The median sequencing depth across the 11 sites was 1,829 reads per position for RTP (IQR 592-3,488) and 2,410 for V3 (IQR 786-3,695). 10 preselected drug resistant variants were measured across sites and showed high inter-laboratory correlation across all sites with data (P0.001). The triplicate samples of a plasmid mixture confirmed the high inter-laboratory consistency (mean% ± stdev: 4.6 ±0.5, 4.8 ±0.4, 4.9 ±0.3) and revealed good intra-laboratory consistency (mean% range ± stdev range: 4.2-5.2 ± 0.04-0.65). In the two dilutions series, no variants >20% were missed, variants 2-10% were detected at most sites (even at low VT), and variants 1-2% were detected by some sites. All mutations detected by population sequencing were also detected by UDS.

Not available in
This is not available in . You can go to the translated versions in these languages: